ISO/IEC 17025:2017

Understanding Decision Rules in ISO 17025:2017

How to account for measurement uncertainty when stating conformity to a specification (Pass/Fail).

Introduction

When a customer asks you to state if a product “Passed” or “Failed” a specification, you are making a statement of conformity. ISO 17025:2017 requires you to document the “decision rule” employed, taking into account the level of risk.

The Problem: Uncertainty

Every measurement has uncertainty. If a limit is 100 and you measure 99 ± 2, is it a Pass?

  • True value could be 101 (Fail).
  • True value could be 97 (Pass).

Types of Decision Rules

1. Simple Acceptance (Shared Risk)

  • Rule: Pass if the measured value is within the limit.
  • Risk: The lab and the customer share the risk of false accept/reject.
  • Usage: Common when uncertainty is small compared to the tolerance.

2. Guard Banding (Stringent Acceptance)

  • Rule: Pass only if the measured value plus the uncertainty is still within the limit.
  • Risk: Reduces risk of “False Accept” (consumer risk) but increases “False Reject” (producer risk).
  • Usage: Critical applications (aerospace, medical).

Communicating with the Customer

You must agree on the decision rule with the customer before starting work (Clause 7.1.3). Don’t surprise them with a “Fail” due to a guard band they didn’t ask for.